Brief communication: Limitations of medical X-ray computed tomography for estimating ice content in permafrost samples

Mahya Roustaei, Jonas Darey, Zakieh Mohammadi, and Daniel Fortier

Viewed

Total article views: 1,884 (including HTML, PDF, and XML)
HTML PDF XML Total BibTeX EndNote
1,186 523 175 1,884 160 159
  • HTML: 1,186
  • PDF: 523
  • XML: 175
  • Total: 1,884
  • BibTeX: 160
  • EndNote: 159
Views and downloads (calculated since 15 Dec 2025)
Cumulative views and downloads (calculated since 15 Dec 2025)

Viewed (geographical distribution)

Total article views: 1,884 (including HTML, PDF, and XML) Thereof 1,865 with geography defined and 19 with unknown origin.
Country # Views %
  • 1
1
 
 
 
 
Latest update: 21 Sep 2026
Download
Short summary
This study assesses the accuracy of medical X-ray computed tomography (CT) for estimating volumetric ice content in permafrost samples. Comparison of 261 cores from northern Canada shows that medical CT systematically underestimates ice in sediment-rich samples and overestimates it in organic-rich ones due to resolution limits and material misclassification. The results highlight medical CT's value for visualizing cryostructures but its limits for quantitative ice estimation.